Latest news, blogs and updates from the children and young people's diabetes team
-
24 August 2023
We have been informed that there has been an increasing number of fractures in the battery securing parts of this pump. During the insulin pump use, if the device is dropped or excessive force is applied, some of the device's battery securing parts may break.
-
24 August 2023
Until this point, if a largely over estimated total daily dose was entered by the user incorrectly, there is a risk of over delivery of insulin during Auto mode. This risk is the highest on the first day and reduces gradually over subsequent days.
-
21 July 2023
We are very excited that the Omnipod 5 has now been launched. The team have been busy over the last few months getting training and ensuring that the pump is on our procurement system.
-
17 July 2023
We have received an Urgent Field Safety Notice regarding the FreeStyle LibreLink app for iOS, version 2.10
-
17 July 2023
We have had many families contacting us recently requesting early delivery of sensors due to going on holiday, or asking when or how to report failed or faulty sensors. It is important to report any issues with sensors, so you do not find yourself short of supplies before the next delivery.